High atrial and ventricular rate episodes in a dual chamber pacemaker were caused by minute ventilation oversensing in the presence of lead connector issues.
Why the study?
Detection of high rate episodes in pacemakers can result from oversensing of physiological or nonphysiological signals, requiring evaluation of lead integrity, lead connection, and electrogram noise characteristics.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Highlights the importance of assessing lead integrity, lead connection, and noise characteristics to diagnose oversensing in pacemakers.
Detection of high rate episodes can potentially result from oversensing of physiological or nonphysiological signals. Assessment of lead integrity, lead connection and analysis of noise characteristics on electrograms are decisive steps in the evaluation of oversensing. We report a case of high atrial and ventricular rate episodes due to minute ventilation oversensing in the presence of lead connector issues.
No takes yet. Share an insight, caveat, or question.
P et al. (2021) conducted a case report in High rate episodes in a dual chamber pacemaker (n=1). Minute ventilation oversensing with lead connector issues was evaluated on High atrial and ventricular rate episodes. High atrial and ventricular rate episodes in a dual chamber pacemaker were caused by minute ventilation oversensing in the presence of lead connector issues.