Key result
A chest radiograph identified excess atrial and ventricular lead slack causing lead-lead interaction and artifacts, which resulted in frequent short episodes of inappropriate mode switching.
Case Report (n=1)
Chest radiography can be a simple and effective tool to identify lead-lead interaction as a cause of inappropriate mode switching in patients with dual-chamber pacemakers.
Answers are grounded in this paper and cite their sources.
May warrant chest radiography for unexplained mode switching in dual-chamber pacemakers; leaves open prospective validation.
An 80-year-old woman with a history of paroxysmal atrial fibrillation and atrioventricular node disease status post-dual chamber pacemaker placement was noted to have abnormal pacing episodes during a percutaneous coronary intervention. Pacemaker interrogation revealed a high number of short duration mode switching episodes. Representative electrograms demonstrated high frequency nonphysiologic recordings predominantly in the atrial lead. Intrinsic pacemaker malfunction was excluded. A chest radiograph showed excess atrial and ventricular lead slack in the right ventricular inflow. It was suspected that lead-lead interaction resulted in artifacts and oversensing, causing frequent short episodes of inappropriate mode switching.
No takes yet. Share an insight, caveat, or question.
Marino et al. (2015) conducted a case report in Inappropriate mode switching (n=1). Chest radiograph was evaluated on Identification of the cause of inappropriate mode switching. A chest radiograph identified excess atrial and ventricular lead slack causing lead-lead interaction and artifacts, which resulted in frequent short episodes of inappropriate mode switching.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: