A loose set screw in a dual chamber pacemaker can cause inappropriate mode switching due to oversensing of a high frequency signal.
Answers are grounded in this paper and cite their sources.
Loose set screw may cause atrial oversensing and inappropriate mode switching; case report leaves open prevalence in device issues.
KURUVILLA, C., et al. : Inappropriate Mode Switching in a Dual Chamber Pacemaker Due to Oversensing of a High Frequency Signal from a Conductor/Ring Discontinuity (Loose Set Screw). Two unusual cases are reported of pacemaker oversensing of the impedance pulse by the atrial channel attributed to a loose set screw.
No takes yet. Share an insight, caveat, or question.
KURUVILLA et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: