Key result
Biotronik pacemakers risk undersensing sustained VT when ventricular rates exceed 240 bpm.
Why the study?
Pacemaker failure to record ventricular high-rate events can result from true or functional undersensing, which requires clarification to improve detection accuracy.
Clinicians must distinguish between true and functional undersensing when pacemakers fail to record ventricular high-rate events, particularly noting device-specific refractory period programming.
Answers are grounded in this paper and cite their sources.
Warrants clinician awareness of Biotronik refractory periods in undersensing cases; leaves open optimal programming to ensure VT detection.
Key Teaching Points•When a pacemaker fails to record a ventricular high-rate (VHR) event, true vs functional undersensing should be considered.•True undersensing refers to failure to detect a VHR event owing to small-amplitude intrinsic signals or the nature of the sensing algorithm.•Functional undersensing occurs if the episode fails to meet the definition of a VHR or is not detected as a result of the way in which refractory or blanking periods are programmed.•Functional undersensing of sustained ventricular tachycardia can occur with Biotronik pacemakers if the ventricular rate exceeds 240 beats per minute owing to the fact that the ventricular refractory period is nominally set to 250 ms and that sensed beats that fall within this refractory period do not increase the VHR counter. •When a pacemaker fails to record a ventricular high-rate (VHR) event, true vs functional undersensing should be considered.•True undersensing refers to failure to detect a VHR event owing to small-amplitude intrinsic signals or the nature of the sensing algorithm.•Functional undersensing occurs if the episode fails to meet the definition of a VHR or is not detected as a result of the way in which refractory or blanking periods are programmed.•Functional undersensing of sustained ventricular tachycardia can occur with Biotronik pacemakers if the ventricular rate exceeds 240 beats per minute owing to the fact that the ventricular refractory period is nominally set to 250 ms and that sensed beats that fall within this refractory period do not increase the VHR counter.
No takes yet. Share an insight, caveat, or question.
Mansour et al. (2018) conducted a review in Ventricular high-rate events. Pacemaker detection of ventricular high-rate events was evaluated. Functional undersensing of sustained ventricular tachycardia can occur with Biotronik pacemakers if the ventricular rate exceeds 240 beats per minute due to a nominal 250 ms refractory period.