Modifying the Biotronik sensing algorithm from a two-point to a single-point registration resolved ventricular undersensing caused by an exceptionally small R wave signal width.
Case Report (n=1)
A specific Biotronik sensing algorithm requiring two consecutive points above maximum sensitivity can cause ventricular undersensing in patients with exceptionally small R wave signal width, which can be resolved by modifying it to single point registration.
Answers are grounded in this paper and cite their sources.
Single-point registration may address undersensing from narrow R waves in Biotronik devices; leaves open prospective validation before wider adoption.
This article reports on a case of ventricular undersensing despite normal R wave amplitudes during sinus rhythm in an ICD patient. Undersensing of ventricular signals was noted without any evidence of lead dislocation or variation in signal amplitude. Undersensing was due to an exceptionally small R wave signal width and a feature of the Biotronik sensing algorithm designed to avoid oversensing. This algorithm, intended to enhance the sensing specificity of the device, requires registration of two consecutive points above the maximum programmed sensitivity for a ventricular sense event. After modifying the algorithm to a single point registration undersensing disappeared.
No takes yet. Share an insight, caveat, or question.
Eberhardt et al. (2003) conducted a case report in Ventricular undersensing in an ICD patient (n=1). Modification of the Biotronik sensing algorithm to a single point registration vs. Two consecutive points registration (original algorithm) was evaluated on Resolution of undersensing. Modifying the Biotronik sensing algorithm from a two-point to a single-point registration resolved ventricular undersensing caused by an exceptionally small R wave signal width.