Why the study?
Does programming the sinus redetection parameter to slow correct undersensing during ventricular tachyarrhythmias in patients with Ventritex Cadence defibrillators?
Population
98 patients with Ventritex Cadence defibrillator systems
Design
Case_series
Follow-up
4-6 weeks following device implant
Authors
Loading...
May avert undersensing-related therapy failure in these ICDs; hypothesis-generating for programming optimization.
Does programming the sinus redetection parameter to slow correct undersensing during ventricular tachyarrhythmias in patients with Ventritex Cadence defibrillators?
Undersensing during ventricular tachyarrhythmias in third-generation ICDs can lead to therapy failure, but can be diagnosed via stored electrograms and corrected by adjusting sinus redetection programming.
Mann et al. (1994) studied this question.