Key result
Decreasing the atrial sensitivity value resolved inappropriate ICD therapy caused by the device misinterpreting supraventricular tachycardia as ventricular tachycardia due to unsensed P waves.
Case Report (n=1)
Adjusting atrial sensitivity can resolve inappropriate ICD therapy caused by atrial undersensing during supraventricular tachycardia.
Answers are grounded in this paper and cite their sources.
May avert inappropriate ICD therapy from atrial undersensing in one case; leaves open generalizability and need for validation.
We present a case of inappropriate detection and therapy in a patient with an implantable cardioverter defibrillator. The device interpreted a supraventricular tachycardia as ventricular tachycardia due to unsensed P waves as a result of a relatively high level of atrial sensitivity setting. Subsequently, inappropriate therapy was delivered. The problem was resolved by decreasing the atrial sensitivity value.
No takes yet. Share an insight, caveat, or question.
Paraskevaidis et al. (2005) conducted a case report in Inappropriate implantable cardioverter defibrillator therapy (n=1). Decreasing the atrial sensitivity value was evaluated on Resolution of inappropriate therapy. Decreasing the atrial sensitivity value resolved inappropriate ICD therapy caused by the device misinterpreting supraventricular tachycardia as ventricular tachycardia due to unsensed P waves.