Key result
Automatic adjustable gain in an implantable cardioverter defibrillator caused inappropriate discharges by falsely sensing atrial tachycardia as ventricular fibrillation after His bundle ablation.
Automatic adjustable gain settings in third-generation ICDs can lead to inappropriate shocks by sensing atrial tachycardia in patients with prior His bundle ablation.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to inappropriate ICD shocks risk post-His ablation; hypothesis-generating for sensing algorithm refinements.
Third generation implantable cardioverter defibrillators are capable of complex arrhythmia detection using sensing algorithms with automatic adjustable gain settings. We report a unique case where inappropriate sensing of atrial tachycardia in a patient with a His bundle ablation lead to satisfaction of ventricular fibrillation detection criteria.