Key result
An implantable cardioverter defibrillator failed to sense every other QRS complex during induced ventricular fibrillation due to marked electrical alternans, despite auto adjustable gain.
Case Report (n=1)
This case highlights that potentially fatal undersensing can occur in ICDs during ventricular fibrillation due to marked electrical alternans, despite the presence of auto adjustable gain.
Answers are grounded in this paper and cite their sources.
May cause ICD undersensing during VF with electrical alternans; leaves open need for sensing algorithm refinements.
WASE A., et.al .: Sensing Failure in a Tiered Therapy Implantable Cardioverter Defibrillator: Role of Auto Adjustable Gain . Implantable cardioverter defibrillators have an established role in the management of life‐threatening tachyarrhythmias. These devices use sophisticated sensing circuitry to detect and promptly treat a vast majority of these arrhythmias. However, they are not foolproof. We report one case where the device failed to sense every other QRS complex during induced ventricular fibrillation due to marked electrical alterans. Thus, undersensing can be a potentially fatal problem despite the use of auto adjustable gain.
No takes yet. Share an insight, caveat, or question.
Wase et al. (1995) conducted a case report in Ventricular fibrillation (n=1). Implantable cardioverter defibrillator with auto adjustable gain was evaluated on Sensing failure. An implantable cardioverter defibrillator failed to sense every other QRS complex during induced ventricular fibrillation due to marked electrical alternans, despite auto adjustable gain.