Key result
An automatic gain feature in a third-generation cardioverter defibrillator caused prolonged ventricular asystole due to far-field sensing of atrial activity and inhibition of ventricular pacing.
Case Report (n=1)
The automatic gain feature in third-generation implantable cardioverter defibrillators can cause prolonged ventricular asystole by inappropriately sensing far-field atrial activity and inhibiting ventricular pacing.
Answers are grounded in this paper and cite their sources.
May warrant vigilance for automatic gain-induced asystole in ICD patients; case report leaves open broader device safety implications.
The diagnostic accuracy of implantable cardioverter defibrillators may be improved by automatically adjusting gain algorithms, which in general reduce the likelihood of oversensing while maintaining the ability to detect the low amplitude signals associated with ventricular fibrillation. We present a patient with a third-generation device who developed prolonged ventricular asystole arising as a complication of the automatic gain feature. During asystole the device automatically increased sensitivity in order to prevent undersensing of ventricular fibrillation, which in this case resulted in far-field sensing of atrial activity and inhibition of ventricular pacing.
No takes yet. Share an insight, caveat, or question.
Curwin et al. (1996) conducted a case report in Patient with a third-generation implantable cardioverter defibrillator (n=1). Automatic gain feature in a third-generation cardioverter defibrillator was evaluated on Prolonged ventricular asystole. An automatic gain feature in a third-generation cardioverter defibrillator caused prolonged ventricular asystole due to far-field sensing of atrial activity and inhibition of ventricular pacing.