Key result
In a patient with long QT syndrome, a third-generation ICD caused inappropriate pauses during bradycardia pacing due to automatic sensitivity increases, which was resolved by noninvasive reprogramming.
Case Report (n=1)
Recognition of specific ICD sensing algorithms like 'lower threshold crossing' is crucial to identify and noninvasively correct inappropriate pacing pauses.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to algorithm-related pacing pauses in LQTS; leaves open prevalence and device-specific prevention strategies.
Many ICD devices have the capability for back up bradycardia pacing. Because of the use of a single sensing algorithm for both bradycardia and tachycardia functions, they may be prone to certain "sensing errors." Following implantation of an ICD in a patient with long QT syndrome, "inappropriate" pauses were noted during bradycardia pacing, which were exactly twice the programmed pacing cycle length. This was due to an automatic increase in the device's sensitivity during pacing, a characteristic of the automatic gain control of this particular ICD. Proper recognition of this ICD's special features, known as "lower threshold crossing," allowed noninvasive rectification of the problem and prevented these pauses.
No takes yet. Share an insight, caveat, or question.
Cossú et al. (1997) conducted a case report in Long QT syndrome (n=1). Third-generation implantable cardioverter defibrillator was evaluated on Inappropriate pauses during bradycardia pacing. In a patient with long QT syndrome, a third-generation ICD caused inappropriate pauses during bradycardia pacing due to automatic sensitivity increases, which was resolved by noninvasive reprogramming.
Synapse has enriched 10 related papers. Consider them for comparative context: