Key result
In a patient with long QT syndrome, a third-generation ICD caused inappropriate pauses during bradycardia pacing due to automatic sensitivity increases, which was resolved by noninvasive reprogramming.
Population
A patient with long QT syndrome following implantation of a third-generation implantable cardioverter…
Design
Case_report
Authors
Loading...
Alerts clinicians to algorithm-related pacing pauses in LQTS; leaves open prevalence and device-specific prevention strategies.
Case Report (n=1)
Recognition of specific ICD sensing algorithms like 'lower threshold crossing' is crucial to identify and noninvasively correct inappropriate pacing pauses.
Cossú et al. (1997) conducted a case report in Long QT syndrome (n=1). Third-generation implantable cardioverter defibrillator was evaluated on Inappropriate pauses during bradycardia pacing. In a patient with long QT syndrome, a third-generation ICD caused inappropriate pauses during bradycardia pacing due to automatic sensitivity increases, which was resolved by noninvasive reprogramming.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: