Key result
Decreasing atrial sensitivity abolishes inappropriate dual-chamber ICD therapies driven by far-field R-wave oversensing.
Why the study?
Far-field oversensing of ventricular beats in the atrial channel of new dual chamber ICDs can cause inappropriate atrial therapies, but detailed descriptions are limited.
Case Report (n=1)
Far-field R wave oversensing in dual chamber ICDs can lead to inappropriate atrial therapies and proarrhythmia, which can be managed by adjusting atrial sensitivity.
Answers are grounded in this paper and cite their sources.
May prompt atrial sensitivity adjustment to avert inappropriate therapies; this case leaves open prospective validation in dual-chamber ICDs.
This case report describes delivery of atrial therapies during a sinus tachycardia in a new dual chamber implantable cardioverter defibrillator inappropriately caused by far-field oversensing of ventricular beats in the atrial channel. Upon classification of the PR interval pattern, the rate criterion for an atrial tachycardia was fulfilled, and the device initiated high-frequency burst pacing as the first stage of programmed tiered atrial therapies. Atrial fibrillation subsequently was induced by high-frequency burst pacing, and eventually a programmed 10-J shock was delivered for successful termination of atrial fibrillation. The phenomenon of far-field oversensing of ventricular beats could be repeatedly observed during exercise testing and abolished by decreasing the atrial sensitivity.
No takes yet. Share an insight, caveat, or question.
Wolpert et al. (1998) conducted a case report in Inappropriate atrial therapies (n=1). Dual chamber implantable cardioverter defibrillator was evaluated. A dual chamber defibrillator delivered inappropriate atrial therapies during sinus tachycardia due to far-field R wave oversensing, which was abolished by decreasing atrial sensitivity.