Electrical Proarrhythmia: Induction of Inappropriate Atrial Therapies due to Far‐Field R Wave Oversensing in a New Dual Chamber Defibrillator

View Full Paper
Ask AI
Bookmark
Share

Key result

Decreasing atrial sensitivity abolishes inappropriate dual-chamber ICD therapies driven by far-field R-wave oversensing.

  • n=1

Why the study?

Far-field oversensing of ventricular beats in the atrial channel of new dual chamber ICDs can cause inappropriate atrial therapies, but detailed descriptions are limited.

Population

A patient with a new dual chamber implantable cardioverter defibrillator

Comparison

Atrial therapies induced by far-field R wave oversensing vs appropriate sensing

Design

Case report

Authors

CWChristian WolpertUniversity of StuttgartWJWerner JungBoston CollegeCSChristopher SchollHeilbronn University

Discussion

Loading...

Member takes

Overview

May prompt atrial sensitivity adjustment to avert inappropriate therapies; this case leaves open prospective validation in dual-chamber ICDs.

Study Design

Type

Case Report (n=1)

Structured PICO

P
Population
A single patient with a new dual chamber implantable cardioverter defibrillator who experienced inappropriate atrial therapies during sinus tachycardia.
I
Intervention
Dual chamber implantable cardioverter defibrillator
O
Outcome
Inappropriate delivery of atrial therapies (high-frequency burst pacing and shock)safety

Far-field R wave oversensing in dual chamber ICDs can lead to inappropriate atrial therapies and proarrhythmia, which can be managed by adjusting atrial sensitivity.

Cite This Study

Wolpert et al. (1998) conducted a case report in Inappropriate atrial therapies (n=1). Dual chamber implantable cardioverter defibrillator was evaluated. A dual chamber defibrillator delivered inappropriate atrial therapies during sinus tachycardia due to far-field R wave oversensing, which was abolished by decreasing atrial sensitivity.

synapsesocial.com/papers/6aa8d6fb552db6915f87259ahttps://doi.org/10.1111/j.1540-8167.1998.tb00126.x
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Value of ventricular electrogram recordings in the diagnosis of arrhythmias precipitating electrical device shock therapy1991 · 141 citations
  2. 2VDD Single Pass Lead Pacing; Sustained Pacemaker Mediated Tachycardias Unrelated to Retrograde Atrial Activation1992 · 19 citations
  3. 3Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human Beings1980 · 1,366 citations
  4. 4Long-term outcome with the automatic implantable cardioverter-defibrillator1989 · 721 citations
  5. 5Incidence of ICD Lead Related Complications During Long‐Term Follow‐Up: Comparison of Epicardial and Endocardial Electrode Systems1995 · 72 citations