Key result
Atrial autoadjusting sensitivity linked to inappropriate AT detection via FFRW oversensing during PR prolongation.
Case Report (n=1)
Demonstrates that far-field R wave oversensing can occur during functional atrial undersensing with PR prolongation in pacemakers with atrial autoadjusting sensitivity, leading to inappropriate AT detection.
Answers are grounded in this paper and cite their sources.
May prompt device interrogation in suspected AT; hypothesis-generating for autoadjusting sensitivity refinements.
COLLINS, R.F., et al.: Rate Dependent Far‐Field R Wave Sensing in an Atrial Tachyarrhythmia Therapy Device. A case is reported of far‐field R wave (FFRW) oversensing resulting in inappropriate atrial tachycardia (AT) detection by a dual chamber pacemaker incorporating atrial autoadjusting sensitivity (AAS). FFRW oversensing occurred during periods of functional atrial undersensing (FAU) with PR interval prolongation. Limitations of the pacemaker's ability to reject FFRWs and programming considerations for addressing this unique behavior are discussed.
No takes yet. Share an insight, caveat, or question.
Collins et al. (2002) conducted a case report in Far-field R wave oversensing and inappropriate atrial tachycardia detection (n=1). Dual chamber pacemaker with atrial autoadjusting sensitivity was evaluated on Inappropriate atrial tachycardia detection. A dual chamber pacemaker with atrial autoadjusting sensitivity caused inappropriate atrial tachycardia detection due to far-field R wave oversensing during PR interval prolongation.