Reprogramming the threshold start can noninvasively correct ICD undersensing caused by ventricular electrogram alternans during ventricular tachycardia.
Answers are grounded in this paper and cite their sources.
May allow noninvasive ICD reprogramming to resolve VEGM undersensing in VT; hypothesis-generating and should not yet change practice.
Alternans of the ventricular electrogram (VEGM) during ventricular tachycardia (VT) is a rare cause of ventricular undersensing by an implantable cardioverter-defibrillator (ICD). This report describes a patient with a St. Jude ICD who exhibited sustained monomorphic VT associated with surface QRS alternans, alternating cycle lengths, alternans of the VEGM causing intermittent undersensing of the smaller component, and intermittent 2:1 counting of ventricular intervals during 1:1 sensing in response to the ICD detection algorithm. VEGM undersensing was corrected noninvasively simply by programming the threshold start from 62.5% to 50% which increased the sensitivity based on the amplitude of the VEGM. This maneuver did not affect the satisfactory and stable defibrillation threshold.