Why the study?
Does the Björk-Shiley prosthetic aortic valve improve hydrodynamic performance compared to other prosthetic valves in an in vitro pulse duplicator system?
Does the Björk-Shiley prosthetic aortic valve improve hydrodynamic performance compared to other prosthetic valves in an in vitro pulse duplicator system?
In an in vitro pulse duplicator system, the Björk-Shiley prosthetic aortic valve demonstrated superior hydrodynamic performance with lower flow resistance, lower energy loss, and predominantly laminar flow compared to four other prosthetic valves.
Answers are grounded in this paper and cite their sources.
Does not justify clinical preference; leaves open translation of hydrodynamic benefits to patient outcomes.
A pulse duplicator system, used for studying prosthetic heart valves, is described. It provides a pulsatile flow resembling that produced by the heart, and permits a quantitative estimation of fluid flow resistance, regurgitation and energy loss in the valves. A technique for visualization of flow pattern is presented. The hydrodynamic performance of five prosthetic aortic valves, Starr-Edwards, Smeloff-Cutter, Kay-Shiley, Wada-Cutter and Björk-Shiley, were studied in the pulse duplicator and compared on the basis of their external (tissue) diameters. For a given tissue diameter, the Björk-Shiley valve had the best performance, resulting in a significantly lower flow resistance and energy loss than in the other prostheses investigated. All valves had acceptably low degrees of regurgitation with the exception of the Wada-Cutter prosthesis (16 to 20% of stroke volume). Flow visualization demonstrated a more or less turbulent flow pattern in all the valves with the exception of the Björk-Shiley valve. In this valve, predominantly laminar flow was observed.
No takes yet. Share an insight, caveat, or question.
Christian Olin (1971) studied this question.