Key result
A novel QRS detection ASIC achieved a sensitivity of 95.65% and a positive prediction of 99.36% on the MIT/BIH arrhythmia database, with a 0.68 mm2 core area and 2.21 μW power consumption.
A newly designed QRS detection ASIC offers high accuracy with minimal power consumption and physical footprint, potentially advancing wearable ECG technology.
May advance low-power wearable ECG devices; leaves open clinical validation.
Healthcare issues arose from population aging. Meanwhile, electrocardiogram (ECG) is a powerful measurement tool. The first step of ECG is to detect QRS complexes. A state‐of‐the‐art QRS detection algorithm was modified and implemented to an application‐specific integrated circuit (ASIC). By the dedicated architecture design, the novel ASIC is proposed with 0.68 mm 2 core area and 2.21 μ W power consumption. It is the smallest QRS detection ASIC based on 0.18 μ m technology. In addition, the sensitivity is 95.65% and the positive prediction of the ASIC is 99.36% based on the MIT/BIH arrhythmia database certification.
No takes yet. Share an insight, caveat, or question.
Huang et al. (2012) studied Arrhythmia. QRS detection ASIC was evaluated on Sensitivity and positive prediction for QRS detection. A novel QRS detection ASIC achieved a sensitivity of 95.65% and a positive prediction of 99.36% on the MIT/BIH arrhythmia database, with a 0.68 mm2 core area and 2.21 μW power consumption.