Key result
An energy-efficient analog electrocardiogram processor achieved an average QRS detection sensitivity of 99.63% and positive predictivity of 99.47% using the MIT-BIH database.
The proposed analog ECG processor achieves high QRS detection sensitivity and positive predictivity with ultra-low power consumption.
An energy-efficient electrocardiogram (ECG) processor for real-time QRS detection is presented. The proposed algorithm is based on the Pan-Tompkins algorithm and it is implemented in the analog domain leveraging ultra-low power analog electronics biased in subthreshold. Operational transconductance amplifiers with ∼100 mV linear range are used in almost all of the processing blocks, while squaring is performed on current signals. Additionally, instead of adaptive thresholding, a fixed-level thresholding is performed, thereby eliminating the need for additional blocks such as memory and threshold update. The processor is designed in 65 nm TSMC CMOS technology and has a footprint of 0.078 mm 2 . When supplied by a 1 V supply, the processor consumes 1.2 nW. Using the recordings in the MIT-BIH database, the processor achieves an average QRS detection sensitivity of 99.63% and positive predictivity of 99.47%.
No takes yet. Share an insight, caveat, or question.
Güngör et al. (2021) studied Electrocardiogram QRS detection. Analog Electrocardiogram Processor was evaluated on Average QRS detection sensitivity and positive predictivity. An energy-efficient analog electrocardiogram processor achieved an average QRS detection sensitivity of 99.63% and positive predictivity of 99.47% using the MIT-BIH database.