Key result
Device extraction relieved constant and intermittent chronic pain in 66% (18 of 27) of patients, though it appeared less successful in women undergoing subsequent reimplantation.
Why the study?
Although unmanageable chronic pain carries a class IIa indication for CIED extraction, limited data exist evaluating causes of pain and the outcomes of extraction in eliminating pain.
Does device extraction relieve chronic pain related to cardiovascular implantable electronic devices?
Observational (n=27)
Yes
Does device extraction relieve chronic pain related to cardiovascular implantable electronic devices?
Device extraction relieves chronic pain at the CIED generator site in about two-thirds of patients, though it appears less successful in women undergoing subsequent reimplantation.
May support extraction for refractory CIED pain; leaves open sex-specific efficacy and reimplantation outcomes.
AIMS: Chronic pain at the cardiovascular implantable electronic device (CIED) generator or lead insertion site that is not otherwise manageable carries a IIA indication for extraction. However, limited data exist evaluating causes of pain and outcomes of extraction in eliminating pain. A multi-centre retrospective observational study was conducted to evaluate outcomes of patients undergoing device extraction for treatment of chronic device pain. METHODS AND RESULTS: Twenty-seven out of 2188 lead extraction candidates (1.3%) met the chronic pain IIA indication for extraction [50 ± 16 years; 14 (51%) women]. Onset, severity, triggers, and pain management were measured before and after extraction. Device type, procedure done (with/without reimplantation), and positive tissue cultures were noted. Pain was reported as constant (n = 14; 50%), intermittent (n = 13; 46%), and movement-triggered (n = 14; 50%). Average severity of pain was seven out of 10 (10 being the worst). Post-extraction, 18 (66%) received freedom from pain, including all patients with poorly formed pockets (n = 2) and subclinical infections (n = 2). Of the 18, 11 underwent reimplantation (61%) without recurrent pain. Nine still had pain (44 ± 17 years; seven women) after extraction. Eight of the nine underwent reimplantation, three on the contralateral chest wall and five ipsilaterally. Pain severity decreased (n = 5), increased (n = 1), or was unchanged (n = 3). CONCLUSION: Chronic pain at the CIED generator site can present as chronic or movement-triggered pain, and can be due to subclinical infection or a poorly formed device pocket. Extraction relieved constant and intermittent pain in two-thirds of patients. Extraction appears less successful in eliminating pain in women who undergo subsequent reimplantation.
No takes yet. Share an insight, caveat, or question.
Lewis et al. (2018) conducted an observational in Chronic pain at the cardiovascular implantable electronic device (CIED) generator or lead insertion site (n=27). Device extraction was evaluated on Freedom from pain post-extraction. Device extraction relieved constant and intermittent chronic pain in 66% (18 of 27) of patients, though it appeared less successful in women undergoing subsequent reimplantation.