Key result
The VmaxST portable metabolic measurement system yielded similar values to the standard Vmax29 system across all measured metabolic and physiological variables (P = 0.88).
Why the study?
Does the VmaxST portable metabolic measurement system provide valid physiological measurements compared to the standard Vmax29 system in asymptomatic males during exercise?
Population
30 asymptomatic, moderately active males
Comparison
VmaxST portable telemetry metabolic measurement… vs Vmax29 standard metabolic measurement system…
Design
Cross-sectional, Tests were performed in random order
Authors
Loading...
Portable systems may suit controlled lab use; leaves open prospective validation across populations and settings.
Cross-Sectional (n=30)
Random order
Does the VmaxST portable metabolic measurement system provide valid physiological measurements compared to the standard Vmax29 system in asymptomatic males during exercise?
p-value: p=0.88
The VmaxST portable metabolic measurement system is a valid device for measuring metabolic and physiological variables during exercise in a controlled laboratory setting.
Laurent et al. (2008) conducted a cross-sectional in Healthy (asymptomatic, moderately active) (n=30). VmaxST portable metabolic measurement system vs. Vmax29 standard measurement system was evaluated on Peak values for time to exhaustion, heart rate, blood pressure, VO2, VCO2, VE, and RER (p=0.88). The VmaxST portable metabolic measurement system yielded similar values to the standard Vmax29 system across all measured metabolic and physiological variables (P = 0.88).