Why the study?
Can echocardiography detect disc dysfunction and leaks in Björk-Shiley tilting disc valve prostheses?
Can echocardiography detect disc dysfunction and leaks in Björk-Shiley tilting disc valve prostheses?
Echocardiography can potentially detect disc dysfunction such as trans-prosthesis leaks in Björk-Shiley valves by identifying reduced motion amplitude and diastolic opening.
Echocardiography may flag Björk-Shiley valve dysfunction in animals; leaves open clinical validation and adoption.
The echograms obtained following Björk-Shiley valve replacement are described-first, on an experimental model and later, in vivo. The echogram obtained consisted essentially of four main horizontal waves. Their origin is clarified. In artificially created peri-prosthesis leaks, a normal trace was obtained, while in trans-prosthesis leaks, a reduction in amplitude of motion occurred and the valve remained open in diastole (aortic valve). In vivo, a normal trace was obtained in peri-prosthesis leaks. Trans-prosthesis leaks with the Björk-Shiley valve have not hitherto been described, but the potential of echoscanning in detecting disc dysfunction is put forward.
No takes yet. Share an insight, caveat, or question.
A. Escarous (1975) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: