Why the study?
Are 3D-based quantitative imaging techniques comparable to 2D-QCA for quantifying luminal dimensions of a bioabsorbable PLLA stent in the acute phase?
Are 3D-based quantitative imaging techniques comparable to 2D-QCA for quantifying luminal dimensions of a bioabsorbable PLLA stent in the acute phase?
Three-dimensional imaging modalities, including non-invasive MSCT, are comparable to 2D-QCA for quantifying luminal dimensions of bioabsorbable PLLA stents in the acute phase.
Answers are grounded in this paper and cite their sources.
2D-QCA may underestimate bioabsorbable stent lengths; leaves open whether 3D modalities improve acute-phase quantification in practice.
AIMS: To investigate if three-dimensional (3D) based quantitative techniques are comparable to each other and to explore possible differences with respect to the reference method of 2D-QCA in the acute phase and to study whether non-invasive MSCT could potentially be applied to quantify luminal dimensions of a stented coronary segment with a novel bioabsorable drug-eluting stent made of poly-l-lactic-acid (PLLA). METHODS AND RESULTS: Quantitative imaging data derived from 16 patients enrolled at our institution in a first-in-man trial (ABSORB) receiving a biodegradable stent and who were imaged with standard coronary angiography and intravascular ultrasound were compared. Shortly, after stenting the patients also underwent a MSCT procedure. Standard 2D-QCA showed significant smaller stent lengths (p < 0.01). Although, the absolute measured stent diameters and areas by 2D-QCA tend to be smaller, the differences failed to be statistically different when compared to the 3D based quantitative modalities. Measurements made by non-invasive QMSCT-CA of implanted PLLA stents appeared to be comparable to the other 3D modalities without significant differences. CONCLUSIONS: Three-dimensional based quantitative analyses showed similar results quantifying luminal dimensions as compared to 2D-QCA during an evaluation of a new bioabsorbable coronary stent design in the acute phase. Furthermore, in biodegradable stents made of PLLA, non-invasive QMSCT-CA can be used to quantify luminal dimensions.
No takes yet. Share an insight, caveat, or question.
Bruining et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: