Key result
Cryoballoon ablation resulted in a 70.9% freedom from atrial fibrillation recurrence at 15 months in patients with persistent AF, compared to 78.9% in patients with paroxysmal AF.
Why the study?
Cryoballoon ablation has become a standard treatment for paroxysmal AF, but limited data are available for outcomes in patients with persistent AF.
Does cryoballoon ablation have different procedural characteristics and recurrence outcomes in patients with persistent versus paroxysmal atrial fibrillation?
Observational (n=944)
Yes
Does cryoballoon ablation have different procedural characteristics and recurrence outcomes in patients with persistent versus paroxysmal atrial fibrillation?
Absolute Event Rate: 70.9% vs 78.9%
p-value: p=0.09
Cryoballoon ablation demonstrates a similar acute efficacy and safety profile in persistent versus paroxysmal atrial fibrillation, despite persistent AF patients having a more diseased substrate.
Supports cryoballoon ablation feasibility in persistent AF; leaves open randomized confirmation of comparable efficacy.
Introduction: Cryoballoon ablation (CBA) has become a standard treatment for paroxysmal atrial fibrillation (PaAF) but limited data is available for outcomes in patients with persistent atrial fibrillation (PeAF). Methods: We analyzed the first 944 patients included in the Spanish Prospective Multi-center Observation Post-market Registry to compare characteristics and outcomes of patients undergoing CBA for PeAF versus PaAF. Results: A total of 944 patients (57.8 ± 10.4 years; 70.1% male) with AF (27.9% persistent) were prospectively included from 25 centers. PeAF patients were more likely to have structural heart disease (67.7 vs. 11.4%; p < 0.001) and left atrium dilation (72.6 vs. 43.3%; p < 0.001). CBA of PeAF was less likely to be performed under general anesthesia (10.7 vs. 22.2%; p < 0.001), with an arterial line (32.2 vs. 44.6%; p < 0.001) and assisted transeptal puncture (11.9 vs. 17.9%; p = 0.025). During an application, PeAF patients had a longer time to −30 °C (35.91 ± 14.20 vs. 34.93 ± 12.87 s; p = 0.021) and a colder balloon nadir temperature during vein isolation (−35.04 ± 9.58 vs. −33.61 ± 10.32 °C; p = 0.004), but received fewer bonus freeze applications (30.7 vs. 41.1%; p < 0.001). There were no differences in acute pulmonary vein isolation and procedure-related complications. Overall, 76.7% of patients were free from AF recurrences at 15-month follow-up (78.9% in PaAF vs. 70.9% in PeAF; p = 0.09). Conclusions: Patients with PeAF have a more diseased substrate, and CBA procedures performed in such patients were more simplified, although longer/colder freeze applications were often applied. The acute efficacy/safety profile of CBA was similar between PaAF and PeAF patients, but long-term results were better in PaAF patients.
No takes yet. Share an insight, caveat, or question.
Vallès et al. (2022) conducted an observational in Atrial Fibrillation (n=944). Persistent atrial fibrillation (undergoing cryoballoon ablation) vs. Paroxysmal atrial fibrillation (undergoing cryoballoon ablation) was evaluated on Freedom from AF recurrences at 15-month follow-up (p=0.09). Cryoballoon ablation resulted in a 70.9% freedom from atrial fibrillation recurrence at 15 months in patients with persistent AF, compared to 78.9% in patients with paroxysmal AF.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: