Key result
Inverting the blood pressure cuff during measurement showed agreement with the standard protocol, with all observed mean differences in systolic and diastolic BP being <5 mmHg across postures.
Why the study?
Accurate and precise measurement is essential for diagnosing hypertension, but whether inverting the cuff bladder causes significant changes in blood pressure readings was unknown.
Does inverting the cuff bladder affect blood pressure readings compared to standard cuff positioning in outpatient individuals?
Cross-Sectional (n=540)
Does inverting the cuff bladder affect blood pressure readings compared to standard cuff positioning in outpatient individuals?
Answers are grounded in this paper and cite their sources.
Inverting the blood pressure cuff bladder does not significantly alter blood pressure readings compared to standard positioning, suggesting the two methods can be used interchangeably.
Introduction: Elevated blood pressure (BP)/hypertension contributes significantly to global morbidity and mortality. It is a silent killer; therefore, earlier diagnosis is imperative. A simple instrument called sphygmomanometer measures BP. Getting an accurate and precise reading is essential for making the diagnosis. Our study sorts to determine if there is any significant change in BP readings when the bladder cuff is inverted during BP measurement. Materials and Methods: This was a cross-sectional study involving 540 individuals who were consecutively recruited from the outpatient clinics over a year period. BP was measured with a mercury sphygmomanometer in three postures of sitting, standing, and supine comparing the standard protocol (control) and our designed protocol of inverted cuff (test). Results: The mean systolic and diastolic BP for all the postures for both sexes were significantly corresponding for both test positions in majority of the age groups; all observed differences in means were <5 mmHg which is within the acceptable limit. The Bland–Altman plot showed significant agreement for the control and test positions in diastole and systole in all the three postures. The mean of the differences of the control and test procedures were −0.38 ± 2.70, 0.27 ± 2.18 (diastole and systole in sitting); −0.46 ± 2.88, 0.09 ± 1.58 (diastole and systole in standing); −0.62 ± 3.14, 0.09 ± 1.82 (diastole and systole in supine). Conclusion: Our test procedure showed agreement with the standard protocol of cuff positioning. Therefore, it could be used interchangeably.
No takes yet. Share an insight, caveat, or question.
Aliyu et al. (2019) conducted a cross-sectional in Blood pressure measurement (n=540). Inverted cuff protocol vs. Standard cuff protocol was evaluated on Mean difference in systolic and diastolic BP. Inverting the blood pressure cuff during measurement showed agreement with the standard protocol, with all observed mean differences in systolic and diastolic BP being <5 mmHg across postures.
Synapse has enriched 10 related papers. Consider them for comparative context: