Key result
Transapical implantation of the Acurate TA device was feasible and safe in 12 animals, with 100% survival and transvalvular mean pressure gradients of 5.4 ± 2.2 mmHg post-deployment.
Why the study?
Does transapical TAVI with the Acurate TA device demonstrate feasibility and safety in swine and sheep models?
Does transapical TAVI with the Acurate TA device demonstrate feasibility and safety in swine and sheep models?
The Acurate TA self-expandable device demonstrated feasibility, safety, and optimal hemodynamic performance in a preclinical animal model of transapical TAVI.
Does not support clinical use of transapical Acurate TA; leaves open human feasibility trials.
OBJECTIVES: Transcatheter aortic valve implantation (TAVI) is currently expanding worldwide, however all available prostheses share some fundamental design drawbacks. We investigated the feasibility, safety and hemodynamic performance of the innovative transapical Acurate TA™ self-expandable device, which has the unique advantage of offering anatomically correct self-alignment within the aortic root. DESIGN: Transapical TAVI was performed in six acute swine and six chronic sheep procedures, with follow-up of 7, 14, 21, 28, 60 and 90 days. TAVI was performed under TEE and angiographic guidance without rapid pacing. RESULTS: A partial sternotomy approach was used to access the LV-apex. All valve implantations were performed as planned and all animals survived the implantation procedure. After deployment, no migration, embolization or coronary obstruction was observed during the observation period. Intraoperative TEE examination identified no signs of intravalvular leakage or valve dysfunction. Transvalvular mean pressure gradients were 5.4 ± 2.2 mmHg decreasing during follow-up (1.6 ± 0.8 mmHg, 1.8 ± 0.8 mmHg, 1.3 ± 0.2, 1.8 ± 0.7 mmHg, 1.6 ± 0.8 mmHg), with a slight increase atday 90 (4.0 ± 2.4 mmHg, P < 0.05). Macroscopic examination at necropsy showed correct anatomical positioning of the valve stent without any signs of structural valve deterioration. CONCLUSIONS: These first results of the innovative self-expandable transapical ACURATE TA™ device explore the feasibility and safety of anatomically correct off-pump implantation with optimal hemodynamic results.
No takes yet. Share an insight, caveat, or question.
Wendt et al. (2012) studied Transcatheter aortic valve implantation (n=12). Acurate TA self-expandable device was evaluated on Feasibility, safety and hemodynamic performance. Transapical implantation of the Acurate TA device was feasible and safe in 12 animals, with 100% survival and transvalvular mean pressure gradients of 5.4 ± 2.2 mmHg post-deployment.