Key result
Assessment of atrial fibrillation burden using an AI-based tool showed high agreement with manual physician assessment, with a calibration slope of 0.975 and bias of -0.006.
Why the study?
High AF burden is associated with adverse outcomes but is not routinely measured in clinical practice, and an AI-based tool could facilitate its assessment.
Does an AI-based tool accurately assess atrial fibrillation burden on 7-day Holter ECGs compared to manual physician assessment in patients with atrial fibrillation?
Cohort (n=82)
Yes
Does an AI-based tool accurately assess atrial fibrillation burden on 7-day Holter ECGs compared to manual physician assessment in patients with atrial fibrillation?
An AI-based tool provides highly accurate assessment of atrial fibrillation burden on 7-day Holter ECGs, demonstrating near-perfect agreement with manual physician assessment.
Answers are grounded in this paper and cite their sources.
Supports AI validation for AF burden on Holter ECGs; leaves open effects on clinical decisions and outcomes.
BackgroundEmerging evidence indicates that a high atrial fibrillation (AF) burden is associated with adverse outcome. However, AF burden is not routinely measured in clinical practice. An artificial intelligence (AI)-based tool could facilitate the assessment of AF burden.ObjectiveWe aimed to compare the assessment of AF burden performed manually by physicians with that measured by an AI-based tool.MethodsWe analyzed 7-day Holter electrocardiogram (ECG) recordings of AF patients included in the prospective, multicenter Swiss-AF Burden cohort study. AF burden was defined as percentage of time in AF, and was assessed manually by physicians and by an AI-based tool (Cardiomatics, Cracow, Poland). We evaluated the agreement between both techniques by means of Pearson correlation coefficient, linear regression model, and Bland-Altman plot.ResultsWe assessed the AF burden in 100 Holter ECG recordings of 82 patients. We identified 53 Holter ECGs with 0% or 100% AF burden, where we found a 100% correlation. For the remaining 47 Holter ECGs with an AF burden between 0.01% and 81.53%, Pearson correlation coefficient was 0.998. The calibration intercept was -0.001 (95% CI -0.008; 0.006), and the calibration slope was 0.975 (95% CI 0.954; 0.995; multiple R2 0.995, residual standard error 0.017). Bland-Altman analysis resulted in a bias of -0.006 (95% limits of agreement -0.042 to 0.030).ConclusionThe assessment of AF burden with an AI-based tool provided very similar results compared to manual assessment. An AI-based tool may therefore be an accurate and efficient option for the assessment of AF burden. Emerging evidence indicates that a high atrial fibrillation (AF) burden is associated with adverse outcome. However, AF burden is not routinely measured in clinical practice. An artificial intelligence (AI)-based tool could facilitate the assessment of AF burden. We aimed to compare the assessment of AF burden performed manually by physicians with that measured by an AI-based tool. We analyzed 7-day Holter electrocardiogram (ECG) recordings of AF patients included in the prospective, multicenter Swiss-AF Burden cohort study. AF burden was defined as percentage of time in AF, and was assessed manually by physicians and by an AI-based tool (Cardiomatics, Cracow, Poland). We evaluated the agreement between both techniques by means of Pearson correlation coefficient, linear regression model, and Bland-Altman plot. We assessed the AF burden in 100 Holter ECG recordings of 82 patients. We identified 53 Holter ECGs with 0% or 100% AF burden, where we found a 100% correlation. For the remaining 47 Holter ECGs with an AF burden between 0.01% and 81.53%, Pearson correlation coefficient was 0.998. The calibration intercept was -0.001 (95% CI -0.008; 0.006), and the calibration slope was 0.975 (95% CI 0.954; 0.995; multiple R2 0.995, residual standard error 0.017). Bland-Altman analysis resulted in a bias of -0.006 (95% limits of agreement -0.042 to 0.030). The assessment of AF burden with an AI-based tool provided very similar results compared to manual assessment. An AI-based tool may therefore be an accurate and efficient option for the assessment of AF burden.
No takes yet. Share an insight, caveat, or question.
Hennings et al. (2023) conducted a cohort in Atrial fibrillation (n=82). AI-based tool (Cardiomatics) vs. Manual assessment by physicians was evaluated on Agreement in AF burden assessment (percentage of time in AF) (Calibration slope 0.975, 95% CI 0.954-0.995). Assessment of atrial fibrillation burden using an AI-based tool showed high agreement with manual physician assessment, with a calibration slope of 0.975 and bias of -0.006.