Key result
Variations in sampling onset caused average greatest shifts in activation time ranging from 11.2 msec at 200 Hz to 0.3 msec at 2,000 Hz, with significant outliers at low sampling rates.
Variations in the initiation of sampling can cause significant outliers in activation time determination, especially at low sampling rates, indicating MP electrograms should be sampled faster than BP electrograms.
May introduce activation time errors at low sampling rates during mapping; leaves open optimal frequency standards for routine electrophysiology use.
The purpose of this work is to determine the sensitivity of the estimated time of peaks and maximum slopes, commonly used in activation time computations, to the instant at which sampling is initiated. Based on complex and quickly changing waveforms, 471 monopolar (MP) and bipolar (BP) epicardial responses in man were selected. These were decimated from 10 kHz to simulate sampling at frequencies ranging from 200 Hz to 2,000 Hz. The peak and maximum absolute slope for BP and the minimum slope for MP were computed repeatedly starting at successive 100 microseconds intervals extending throughout the sampling period and compared with these parameters computed from the waveform sampled at 10 kHz. Slopes were estimated using each of four different algorithms. The average greatest shift (AGS) due to variations in sampling onset ranged from 11.2 +/- 3.5 (200 Hz) to 0.3 +/- 0.2 msec (2,000 Hz). For bipolar algorithms, the peak performed better than the slope algorithms (AGS: 5.9 +/- 3.3 to 0.3 +/- 1.0 msec). For MP algorithms, 2 point linear, and 3 and 5 point Lagrange slope estimates performed similarly (AGS: 5.6 +/- 3.3 to 0.3 +/- 0.2 msec); a 5 point least square fit algorithm performed poorly. Sampling MP and BP electrograms below 500 and 400, respectively, often caused maximum shifts greater than 4 msec. Thus, the resolution of the peak and estimated slope is not limited to the sampling period, variations in initiation of sampling can cause significant outliers especially at low sampling rates, and MP electrograms should be sampled faster than BP electrograms for comparable accuracy.
No takes yet. Share an insight, caveat, or question.
Pieper et al. (1991) studied Intraoperative mapping. Variations in sampling onset and frequency vs. 10 kHz sampling frequency was evaluated on Average greatest shift (AGS) in estimated time of peaks and maximum slopes. Variations in sampling onset caused average greatest shifts in activation time ranging from 11.2 msec at 200 Hz to 0.3 msec at 2,000 Hz, with significant outliers at low sampling rates.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: