Key result
A novel 3-D high frame-rate ultrasound framework showed good agreement with an external flow rate meter on phantom recordings and clinical feasibility in a single patient.
Why the study?
Multiparametric echocardiographic assessment of valve regurgitation is time-consuming and examiner-dependent, motivating quantitative markers to facilitate evaluation.
A novel 3-D high frame-rate ultrasound framework demonstrates accurate quantification of regurgitant flow rates and volumes in simulations and phantoms, with initial clinical feasibility shown in a patient with mitral regurgitation.
Answers are grounded in this paper and cite their sources.
Supports further 3-D ultrasound development for regurgitant flow; leaves open validation in larger patient cohorts.
Valve regurgitation is a cardiac condition caused by the incomplete closure of a cardiac valve. Untreated, this condition may result in cardiac failure. Regular monitoring of this condition is essential in guiding the decision process for surgical intervention. Current guidelines recommend a multiparametric assessment of valve regurgitation using echocardiography, which is both time consuming and heavily dependent on the experience of the examiner. Several methods have been proposed to provide quantitative markers to facilitate the assessment of valve regurgitation, most notably the Proximal Isovelocity Surface Area (PISA) method and methods based on the quantification of the total Regurgitant Volume (RVol) from the power of backscattered blood signal. In this work, we propose a framework based on trans-thoracic 3-D high frame-rate acquisitions for the simultaneous estimation of the jet cross-sectional area and jet velocity directly at the jet core, which are then combined to estimate the instantaneous flow rate and RVol patients with aortic or mitral insufficiency. We compare two methods for the segmentation of the jet cross-sectional area from the power Doppler signal. Validation on simulated data indicates good segmentation accuracy for the best method (β = 0.97, R 2 = 0.91). Validation on recordings from a flow phantom shows good agreement (β = 1.2, R 2 = 0.88) with an external flow rate meter. Clinical feasibility of the method is also shown in a patient with mitral regurgitation.
No takes yet. Share an insight, caveat, or question.
Fiorentini et al. (2023) studied Valve regurgitation (n=1). Trans-thoracic 3-D high frame-rate ultrasound framework vs. External flow rate meter (in phantom validation) was evaluated on Segmentation accuracy and flow rate agreement. A novel 3-D high frame-rate ultrasound framework showed good agreement with an external flow rate meter on phantom recordings and clinical feasibility in a single patient.
Synapse has enriched 10 related papers. Consider them for comparative context: