Population
94-year-old man with a prior history of coronary artery disease presenting with altered mental status
Comparison
Testing with an alternative troponin assay vs Initial troponin assay
Design
Case_report
Key result
Immunological interference due to heterophile antibodies caused falsely elevated cardiac troponin I levels on the AccuTnI+3 assay, which was undetectable on the Elecsys assay.
Authors
Loading...
Consider assay interference with discordant troponin results; this case leaves open assay standardization needs.
Case Report (n=1)
Discordance between clinical presentation and elevated troponin levels should prompt investigation for false-positive results due to assay interference, such as heterophile antibodies.
Ayan et al. (2018) conducted a case report in Falsely elevated cardiac troponin (n=1). Alternative troponin assay (Elecsys Troponin I Assay, Roche) vs. Locally used assay (AccuTnI+3, Beckman Coulter) was evaluated on Cardiac troponin level. Immunological interference due to heterophile antibodies caused falsely elevated cardiac troponin I levels on the AccuTnI+3 assay, which was undetectable on the Elecsys assay.