Key result
An Apple smartwatch failed to detect ventricular tachycardia with a heart rate over 200 beats/min in a 36-year-old woman with arrhythmogenic right ventricular dysplasia.
Case Report (n=1)
No
Smartwatches relying on photoplethysmography may fail to detect significant tachyarrhythmias such as ventricular tachycardia, highlighting the danger of relying solely on these devices for ambulatory arrhythmia monitoring.
Answers are grounded in this paper and cite their sources.
May miss high-rate VT; case report leaves open smartwatch reliability for arrhythmia monitoring in ARVD.
Arrhythmogenic right ventricular dysplasia (ARVD) is caused by mutations in genes coding for components of desmosomes in the myocardium. Mutations in these genes make desmosomes dysfunctional and account for myocyte detachment, followed by inflammation and apoptosis when it encounters undue mechanical stress. This is why ARVD is a common cause of sudden cardiac death in athletes with undiagnosed ARVD, as increased physical activity exacerbates this progression of ARVD and associated arrhythmias. We describe a case of ARVD in a 36-year-old woman who presented with an unusual sensation in her chest due to non-sustaining ventricular tachycardia, which her smartwatch failed to pick up. Many smartwatches use photoplethysmography (PPG) to monitor heart rate (HR). A typical PPG device contains two light sources (green light and infrared) and a photodetector to measure the reflected light, proportional to the beat-to-beat variation in blood volume. HR is then calculated from these variations. In ambulatory settings, smartwatches underestimate HR in most tachyarrhythmias, mainly when the HR is more than 100 beats/min. Patients using smartwatches for ambulatory heart monitoring should know that the absence of an irregular pulse notification does not exclude possible arrhythmias. Management of ARVD is mainly focused on the prevention of syncope and cardiac arrest through antiarrhythmic medications and an implantable cardioverter defibrillator.
No takes yet. Share an insight, caveat, or question.
Sivakumar et al. (2021) conducted a case report in Arrhythmogenic right ventricular dysplasia (ARVD) (n=1). Smartwatch (Apple Watch Series 3) vs. Pulse oximeter / ECG was evaluated on Detection of ventricular tachycardia. An Apple smartwatch failed to detect ventricular tachycardia with a heart rate over 200 beats/min in a 36-year-old woman with arrhythmogenic right ventricular dysplasia.