Answers are grounded in this paper and cite their sources.
This report demonstrates a method to quantify stenosis and detect plaques in angiograms, indicating a new diagnostic approach.
This report describes a procedure to quantify stenosis and to detect enface plaques in single plane angiograms. We derived an optical density function proportional to local arterial chord length from arteriographs over a wide range of conditions. Standard radiographs were made of silicone rubber bolus used as the radiopaque medium in cadaver arteries. Replicate arterial casts filled with iodine-bearing contrast medium were radiographed for comparison. Identical atherosclerotic areas in both film sets were scanned with a microdensitometer and vessel chord lengths used to construct arterial cross sections for comparison with vessel casts. Cross-sectional areas derived from cadaver and phantom films were highly correlated with those derived from cast slices (r=0.997). Thus, luminal cross-sectional areas were accurately estimated independent of plaque location, and accuracy of chord length measurement was ± 0.28 mm in vessels four to 10 mm in diameter.
No takes yet. Share an insight, caveat, or question.
Crawford et al. (1977) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: