Key result
Attempted implantation of nonthoracotomy defibrillation lead systems using monophasic shock waveforms had a higher failure rate than biphasic waveforms.
Why the study?
Does the use of biphasic shock waveforms reduce implantation failure of nonthoracotomy defibrillation lead systems compared to monophasic waveforms?
Cohort (n=243)
Does the use of biphasic shock waveforms reduce implantation failure of nonthoracotomy defibrillation lead systems compared to monophasic waveforms?
The use of biphasic shock waveforms markedly improves the success rate of nonthoracotomy defibrillation lead system implantation by reducing the incidence of unacceptably high defibrillation thresholds.
Answers are grounded in this paper and cite their sources.
Biphasic waveforms were associated with lower nonthoracotomy lead implantation failure; leaves open need for randomized confirmation before practice change.
Two hundred forty-three consecutive patients underwent attempted implantation of nonthoracotomy defibrillation lead (NTL) systems. The importance of clinical and lead-related factors were analyzed regarding their relation with implantation failure caused by an unacceptably high defibrillation threshold (DFT). Overall, 33 (14%) of 243 patients failed NTL implantation. Patients undergoing attempted implantation of NTL systems with monophasic shock waveforms (monophasic group, n = 145) had an incidence of failed implantation of 22% (n = 32) versus an incidence of 1% (n = 1) among patients undergoing attempted implantation by using biphasic shock waveforms (biphasic group, n = 98; odds ratio, 26.9; p < 0.001). The incidence of success and simplicity of implantation of NTL systems was markedly improved in patients undergoing NTL implantation by using biphasic shock waveforms. Clinical factors could be used to stratify patients in the monophasic group for their risk of implantation failure. In the biphasic group, no clinical factor could be correlated with a low DFT with a fully endovascular system.
No takes yet. Share an insight, caveat, or question.
Schwartzman et al. (1996) conducted a cohort in Attempted implantation of nonthoracotomy defibrillation lead systems (n=243). Monophasic shock waveforms vs. Biphasic shock waveforms was evaluated on Implantation failure caused by an unacceptably high defibrillation threshold (OR 26.9, p=<0.001). Attempted implantation of nonthoracotomy defibrillation lead systems using monophasic shock waveforms had a higher failure rate than biphasic waveforms.
Synapse has enriched 10 related papers. Consider them for comparative context: