Key result
In patients with a multilead nonthoracotomy ICD system, the defibrillation threshold significantly increased 2-8 days postoperatively (22.7 J vs 16.9 J at implant; P<0.05) before returning to baseline.
Why the study?
Does the defibrillation threshold change early postoperatively compared to implant in patients receiving a nonthoracotomy ICD system?
Population
15 patients undergoing implantation of an implantable cardioverter defibrillator, predominantly using a…
Comparison
Defibrillation threshold testing at 2-8 days… vs Defibrillation threshold at the time of implant.
Design
Cohort
Follow-up
4-6 weeks
Authors
Loading...
May warrant early postoperative DFT vigilance in nonthoracotomy ICDs; leaves open validation in contemporary systems and outcome impact.
Cohort (n=15)
Does the defibrillation threshold change early postoperatively compared to implant in patients receiving a nonthoracotomy ICD system?
Absolute Event Rate: 22.7% vs 16.9%
p-value: p=<0.05
Defibrillation thresholds transiently increase early after nonthoracotomy ICD implantation before returning to baseline by 4-6 weeks, which has important implications for programming energy output.
Hsia et al. (1994) conducted a cohort in Implantable cardioverter defibrillator implantation (n=15). Implantable cardioverter defibrillator implantation vs. Baseline (at implant) was evaluated on Defibrillation threshold (DFT) at 2-8 days postoperation (p=<0.05). In patients with a multilead nonthoracotomy ICD system, the defibrillation threshold significantly increased 2-8 days postoperatively (22.7 J vs 16.9 J at implant; P<0.05) before returning to baseline.