Key result
In patients with a multilead nonthoracotomy ICD system, the defibrillation threshold increased 2-8 days postoperatively before returning to baseline.
Why the study?
Does the defibrillation threshold change early postoperatively compared to implant in patients receiving a nonthoracotomy ICD system?
Cohort (n=15)
Does the defibrillation threshold change early postoperatively compared to implant in patients receiving a nonthoracotomy ICD system?
Defibrillation thresholds transiently increase early after nonthoracotomy ICD implantation before returning to baseline by 4-6 weeks, which has important implications for programming energy output.
Answers are grounded in this paper and cite their sources.
May warrant early postoperative DFT vigilance in nonthoracotomy ICDs; leaves open validation in contemporary systems and outcome impact.
The stability of the defibrillation threshold (DFT) early after implantation of an implantable cardioverter defibrillator was evaluated in 15 patients. All but one patient had a three lead nonthoracotomy system using a subcutaneous patch, a right ventricular endocardial lead, and a lead in coronary sinus (n = 5) or superior vena cava (n = 9). Shocks were delivered using simultaneous in nine, sequential in three, and single pathway (coronary sinus not used) in one patient. DFTs were measured at implant (n = 15), 2-8 days postoperation (postop, n = 15), and 4-6 weeks later (n = 8). The DFT was defined as the lowest energy shock that resulted in successful defibrillation. The DFT was assessed with output beginning at 18 joules or 2-4 joules above the implant DFT. All shocks were delivered in 2- to 4-joule increments or decrements. DFTs were significantly higher postoperatively than DFTs at implant (22.7 +/- 7.0 J vs 16.9 +/- 3.9 J; P < 0.05). Eight of 15 patients had DFT determined at all three study periods. In these patients, DFT increased at postop (22.8 +/- 8.3 J vs 16.4 +/- 3.9 J at implant; P < 0.05) and returned to baseline at 4-6 weeks (16 +/- 7.1 J vs 16.4 +/- 3.9 J at implant; P = N.S.). Thus, in patients with a multilead nonthoracotomy system, a DFT rise was observed early after implant. The DFT appears to return to baseline in 4-6 weeks. These results have important implications for programming energy output after implantable cardioverter defibrillator implantation.
No takes yet. Share an insight, caveat, or question.
Hsia et al. (1994) conducted a cohort in Implantable cardioverter defibrillator implantation (n=15). Implantable cardioverter defibrillator implantation vs. Baseline (at implant) was evaluated on Defibrillation threshold (DFT) at 2-8 days postoperation (p=<0.05). In patients with a multilead nonthoracotomy ICD system, the defibrillation threshold increased 2-8 days postoperatively before returning to baseline.
Synapse has enriched 10 related papers. Consider them for comparative context: