Key result
Upright position linked to ~1.9-J higher defibrillation threshold vs supine.
Why the study?
Defibrillation thresholds are usually determined in the supine position, but patients may be upright during shocks, potentially explaining some first shock failures.
Does upright body position increase defibrillation thresholds in patients with nonthoracotomy implantable defibrillators compared to the supine position?
Observational (n=52)
Does upright body position increase defibrillation thresholds in patients with nonthoracotomy implantable defibrillators compared to the supine position?
Defibrillation thresholds for biphasic shocks are significantly higher in the upright position compared to the supine position, suggesting first shock energy should be programmed to account for this difference.
Answers are grounded in this paper and cite their sources.
Upright position associated with higher DFT; hypothesis-generating for position-specific programming, needs prospective confirmation.
DFT of Nonthoracotomy Defibrillators. Introduction: Defibrillation thresholds (DFTs) usually are determined with the patient in the supine position. However, patients may be in the upright position when a shock is delivered during follow‐up, which may explain some first shock failures observed clinically. This study investigated whether body posture affects defibrillation energy requirements of nonthoracotomy implantable cardioverter defibrillators with biphasic shocks. Methods and Results: Using a step up‐down protocol, DFTs were compared intraindividually in 52 patients (“active‐can” systems in 41 patients, two‐lead systems in II patients) for the supine and upright positions as achieved by a tilt table. The mean DFT was 7.3 ± 4.2 J in the supine versus 9.2 ± 4.8 J in the upright position (P = 0.002). Repeated comparison in reversed order 3 months after implantation in 22 patients revealed thresholds of 6.2 ± 2.5 J (supine) versus 8.4 ± 3.7 J (upright; P < 0.03) 1 week and 4.4 ± 2.4 J (supine) versus 6.2 ± 4.1 J (upright; P < 0.04) 3 months after implantation. DFTs decreased significantly for both body positions from 1 week to 3 months after implantation (P < 0.04). Conclusion:(I) DFTs for biphasic shocks delivered by nonthoracotomy defibrillators are higher in the upright compared to the supine body position. (2) Differences remain significant 3 months after implantation. For both body positions, DFT decreases significantly from 1 week to 3 months after implantation. These findings have important implications for programming first shock energy to lower than maximal values or for development of devices with lower maximal stored energy.
No takes yet. Share an insight, caveat, or question.
Schauerte et al. (1998) conducted an observational in Implantable cardioverter defibrillator patients (n=52). Upright body position vs. Supine body position was evaluated on Mean defibrillation threshold (DFT) energy requirement (p=0.002). The mean defibrillation threshold for biphasic shocks was higher in the upright body position compared to the supine position.
Synapse has enriched 10 related papers. Consider them for comparative context: