Key result
Addition of an azygos vein coil successfully reduced the defibrillation threshold and allowed successful defibrillation at 15 J in the upright position after the original configuration failed.
Why the study?
Does an upright position increase defibrillation threshold and lead to defibrillation failure in patients with an ICD?
Case Report (n=1)
No
Does an upright position increase defibrillation threshold and lead to defibrillation failure in patients with an ICD?
Answers are grounded in this paper and cite their sources.
Defibrillation threshold may increase significantly in the upright position, potentially leading to defibrillation failure even if supine testing was successful.
It has previously been reported that patients with an implantable cardioverter-defibrillator (ICD) may have a slightly higher defibrillation threshold (DFT) in an upright position compared to a supine position.1 However, the significance of this and the potential impact on clinical practice remains unclear. We report a case of defibrillation failure caused by an increase in DFT in the upright position, leading to prolonged, near-fatal cardiac arrest in a patient who had previous ICD implantation including successful DFT testing in a supine position.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2016) conducted a case report in Nonischemic dilated cardiomyopathy with ICD (n=1). ICD system revision with azygos vein coil vs. Original single-coil ICD configuration was evaluated on Defibrillation threshold in upright position. Addition of an azygos vein coil successfully reduced the defibrillation threshold and allowed successful defibrillation at 15 J in the upright position after the original configuration failed.
Synapse has enriched 10 related papers. Consider them for comparative context: