Key result
Right ventriculography guided lead anchoring to the right ventricular septum in 98% of patients, and mid-RVS pacing yielded a narrower QRS duration than other RVS sites.
Why the study?
Does right ventriculography aid in accurately anchoring a pacing lead to the right ventricular septum in patients undergoing pacemaker implantation?
Observational (n=84)
Does right ventriculography aid in accurately anchoring a pacing lead to the right ventricular septum in patients undergoing pacemaker implantation?
Right ventriculography is a highly effective method for guiding accurate pacing lead placement in the right ventricular septum, with mid-septal pacing yielding the narrowest QRS duration.
Answers are grounded in this paper and cite their sources.
Right ventriculography may facilitate RVS lead placement; observational data leave open whether mid-septal positioning improves outcomes.
Aims: Although right ventricular septal pacing is thought to be more effective in minimizing pacing-induced left ventricular dysfunction, the accurate way to anchor the lead to the right ventricular septum (RVS) has not been established. Our aim was to clarify the usefulness of right ventriculography (RVG) to aid accurate anchoring of the lead to the RVS. Methods and results: Eighty-four patients who underwent pacemaker implantation were enrolled. We anchored the lead to the RVS by using an RVG image obtained at a 30° right anterior oblique view as a reference. We confirmed the actual lead position by performing computed tomography after the procedure and examined the characteristics of the paced QRS complex. Of the 81 patients, except 3 patients whose leads were anchored to the apex due to high pacing thresholds in the RVS, the leads were successfully anchored to the RVS in the 79 (98%) patients, and the number of leads placed in the high-, mid-, and low-RVS was 3 (4%), 58 (73%), and 18 (23%), respectively. The paced QRS duration in these 79 patients was 140 ± 13 ms. The paced QRS duration from mid-RVS was considerably narrower than that from high- or low-RVS (137 ± 12 ms vs. 146 ± 12 ms; P = 0.012). Conclusion: Right ventriculography was very useful in aiding accurate anchoring of the lead to the RVS. Further, pacing from mid-RVS may be more effective in minimizing the QRS duration than pacing from other RVS sites.
No takes yet. Share an insight, caveat, or question.
Shimeno et al. (2017) conducted an observational in Pacemaker implantation (n=84). Right ventriculography-guided lead anchoring to the mid-right ventricular septum vs. Anchoring to high- or low-right ventricular septum was evaluated on Paced QRS duration (p=0.012). Right ventriculography guided lead anchoring to the right ventricular septum in 98% of patients, and mid-RVS pacing yielded a narrower QRS duration than other RVS sites.