Key result
Nexfin noninvasive arterial pressure monitoring provided accurate blood pressure measurements compared to auscultation, with a median systolic difference of 5.4 mm Hg and diastolic of -2.5 mm Hg.
Why the study?
Does the Nexfin noninvasive continuous blood pressure monitor provide accurate measurements compared to standard auscultatory Riva-Rocci/Korotkoff measurements in supine subjects?
Observational (n=104)
Does the Nexfin noninvasive continuous blood pressure monitor provide accurate measurements compared to standard auscultatory Riva-Rocci/Korotkoff measurements in supine subjects?
Effect estimate: Median difference 5.4 mm Hg (systolic) and -2.5 mm Hg (diastolic)
The Nexfin monitor provides accurate noninvasive continuous blood pressure measurements with good within-subject precision compared to standard auscultatory methods.
May support Nexfin for continuous BP monitoring in supine patients; leaves open prospective validation before routine use.
BACKGROUND: The Finapres methodology offers continuous measurement of blood pressure (BP) in a noninvasive manner. The latest development using this methodology is the Nexfin monitor. The present study evaluated the accuracy of Nexfin noninvasive arterial pressure (NAP) compared with auscultatory BP measurements (Riva-Rocci/Korotkoff, RRK). METHODS: In supine subjects NAP was compared to RRK, performed by two observers using an electronic stethoscope with double earpieces. Per subject, three NAP-RRK differences were determined for systolic and diastolic BP, and bias and precision of differences were expressed as median (25th, 75th percentiles). Within-subject precision was defined as the (25th, 75th percentiles) after removing the average individual difference. RESULTS: A total of 312 data sets of NAP and RRK for systolic and diastolic BP from 104 subjects (aged 18-95 years, 54 males) were compared. RRK systolic BP was 129 (115, 150), and diastolic BP was 80 (72, 89), NAP-RRK differences were 5.4 (-1.7, 11.0) mm Hg and -2.5 (-7.6, 2.3) mm Hg for systolic and diastolic BP, respectively; within-subject precisions were (-2.2, 2.3) and (-1.6, 1.5) mm Hg, respectively. CONCLUSION: Nexfin provides accurate measurement of BP with good within-subject precision when compared to RRK.
No takes yet. Share an insight, caveat, or question.
Schattenkerk et al. (2009) reported an observational. Nexfin noninvasive arterial pressure (NAP) vs. Auscultatory BP measurements (Riva-Rocci/Korotkoff, RRK) was evaluated on Difference between NAP and RRK for systolic and diastolic BP (Median difference 5.4 mm Hg (systolic) and -2.5 mm Hg (diastolic)). Nexfin noninvasive arterial pressure monitoring provided accurate blood pressure measurements compared to auscultation, with a median systolic difference of 5.4 mm Hg and diastolic of -2.5 mm Hg.