Key result
Non-invasive pulse pressure variation measured with Nexfin differed from radial artery catheter measurements.
Why the study?
Does the non-invasive Nexfin device accurately measure arterial pressure and pulse pressure variation compared to a radial artery catheter in patients following major upper abdominal surgery?
Observational (n=19)
Does the non-invasive Nexfin device accurately measure arterial pressure and pulse pressure variation compared to a radial artery catheter in patients following major upper abdominal surgery?
The Nexfin monitor system provides acceptable non-invasive arterial pressure monitoring in patients after major upper abdominal surgery without major cardiovascular compromise.
Answers are grounded in this paper and cite their sources.
Nexfin PPV differs significantly from radial artery measurements; leaves open reliability of non-invasive monitoring after major upper abdominal surgery.
We compared the accuracy and precision of the non-invasive Nexfin(®) device for determining systolic, diastolic, mean arterial pressure and pulse pressure variation, with arterial blood pressure values measured from a radial artery catheter in 19 patients following upper abdominal surgery. Measurements were taken at baseline and following fluid loading. Pooled data results of the arterial blood pressures showed no difference between the two measurement modalities. Bland-Altman analysis of pulse pressure variation showed significant differences between values obtained from the radial artery catheter and Nexfin finger cuff technology (mean (SD) 1.49 (2.09)%, p < 0.001, coefficient of variation 24%, limits of agreement -2.71% to 5.69%). The effect of volume expansion on pulse pressure variation was identical between methods (concordance correlation coefficient 0.848). We consider the Nexfin monitor system to be acceptable for use in patients after major upper abdominal surgery without major cardiovascular compromise or haemodynamic support.
No takes yet. Share an insight, caveat, or question.
Wilde et al. (2016) conducted an observational in major upper abdominal surgery (n=19). Nexfin device vs. radial artery catheter was evaluated on pulse pressure variation (MD 1.49%, 95% CI -2.71 to 5.69, p=< 0.001). Non-invasive pulse pressure variation measured with Nexfin differed from radial artery catheter measurements.