Key result
A new multimode transreflectance PPG sensor had a 2.4% failure rate in estimating SpO2 during induced hypothermia (13-21 °C), compared to 70% for a commercial pulse oximeter.
Why the study?
Does a new multimode transreflectance PPG sensor reduce the failure rate of SpO2 estimation compared to a commercial pulse oximeter during induced hypothermia in healthy volunteers?
Does a new multimode transreflectance PPG sensor reduce the failure rate of SpO2 estimation compared to a commercial pulse oximeter during induced hypothermia in healthy volunteers?
Absolute Event Rate: 2.4% vs 70%
A newly developed multimode transreflectance PPG sensor significantly reduces the failure rate of SpO2 estimation during induced hypothermia compared to standard commercial pulse oximeters.
May enable reliable SpO2 monitoring in hypothermia; hypothesis-generating pending validation in patients.
Pulse oximetry utilizes the technique of photoplethysmography to estimate arterial oxygen saturation (SpO(2)) values. During hypothermia, the amplitude of the photoplethysmograph (PPG) is compromised which can lead to inaccurate estimation of SpO(2). A new mutlimode PPG/pulse oximeter sensor was developed to investigate the behaviour of PPGs during conditions of induced hypothermia (hand immersed in an ice bath). PPG measurements from 20 volunteers were conducted and SpO(2) values were estimated at all stages of the experiment. Good quality PPG signals were observed from the majority of the volunteers at almost all hand temperatures. At low temperature ranges, from 13 to 21 °C, the failure rate to estimate SpO(2) values from the multimode transreflectance PPG sensor was 2.4% as compared to the commercial pulse oximeter with a failure rate of 70%.
No takes yet. Share an insight, caveat, or question.
Shafique et al. (2012) studied Healthy volunteers (n=20). Multimode transreflectance PPG sensor vs. Commercial pulse oximeter was evaluated on Failure rate to estimate SpO2 values at low temperature ranges (13 to 21 °C). A new multimode transreflectance PPG sensor had a 2.4% failure rate in estimating SpO2 during induced hypothermia (13-21 °C), compared to 70% for a commercial pulse oximeter.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: