Key result
BioPoint wearable accurately measures heart rate and oxygen saturation from various body locations versus medical-grade oximetry.
Why the study?
An in-depth assessment of the biophysical signals recorded by the BioPoint wearable device, as well as its ability to derive vital signs, remained to be carried out.
Does the BioPoint wearable device accurately measure heart rate and oxygen saturation compared to a medical-grade pulse oximeter?
Observational
Does the BioPoint wearable device accurately measure heart rate and oxygen saturation compared to a medical-grade pulse oximeter?
Answers are grounded in this paper and cite their sources.
The BioPoint wearable device can accurately measure heart rate and oxygen saturation from multiple body sites, but requires calibration before it can be used for medical purposes.
The BioPoint is a new wireless and wearable device, targeting both the ambulatory and on-site monitoring of biosignals. It is described as being capable of streaming and recording the i) electromyography, ii) electrocardiography, iii) electrodermal activity, iv) photoplethysmography, v) skin temperature and vi) actigraphy simultaneously, while making the raw signals recorded by the sensors readily available. However, an in-depth assessment of the biophysical signals recorded by this device, as well as its ability to derive vital signs and other health metrics, remains to be carried out. Consequently, this work proposes a preliminary study to evaluate the quality of the signals that can be acquired by this wearable with a focus on the derivation of heart rate and peripheral blood oxygenation via photoplethysmography. The device is quantitatively compared to the medical-grade pulse oximeter NoninConnect 3245, by Nonin inc. This study was performed with participants wearing the BioPoint at different positions on the body (finger, wrist, forearm, biceps and plantar arch), while the NoninConnect was worn on the fingertip and used as the ground truth. The results show that the BioPoint can accurately determine both heart rate and oxygen saturation from various locations on the body. However, as the BioPoint's photoplethysmograph is not calibrated it cannot be used for medical purposes (non-medical-grade).
No takes yet. Share an insight, caveat, or question.
Gagnon-Turcotte et al. (2023) reported an observational. BioPoint wearable device vs. NoninConnect 3245 pulse oximeter was evaluated on Heart rate and peripheral blood oxygenation via photoplethysmography. The BioPoint wearable device accurately determined heart rate and oxygen saturation from various body locations compared to a medical-grade pulse oximeter.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: