Key result
The tri-wavelength MAX30101 sensor placed on the forehead was more accurate in measuring heart rate, blood oxygen saturation, and respiration rate compared to the bi-wavelength MAX30102 sensor.
Why the study?
Off-the-shelf photoplethysmography sensors in wearable-friendly form factors are increasingly available, prompting evaluation of tri-wavelength and bi-wavelength forehead sensors against an approved reference device.
Do forehead-placed bi-wavelength and tri-wavelength photoplethysmography sensors accurately measure vital signs compared to a standard finger device?
Cross-Sectional
Do forehead-placed bi-wavelength and tri-wavelength photoplethysmography sensors accurately measure vital signs compared to a standard finger device?
Answers are grounded in this paper and cite their sources.
The tri-wavelength MAX30101 forehead sensor provides more accurate vital sign measurements than the bi-wavelength MAX30102 sensor when compared to a standard finger PPG device.
Photoplethysmography is a well-established minimally invasive method for estimating heart rate and blood oxygen saturation in clinical settings. Photoplethysmography, with limitations, can also be extended to estimate respiration rate. Recently many off the shelf photoplethysmography sensors have been released in a form factor suitable for wearable devices. In this study we evaluate two photoplethysmography front ends (inclusive of suitable signal conditioning module and digitizer) from Maxim Integrated™. The tri-wavelength MAX30101 sensor and bi-wavelength MAX30102 sensor were evaluated on the forehead, in a truly non-obtrusive wearable friendly position. Heart rate, respiration rate and blood oxygen saturation were extracted from both sensors and compared with a FDA/TGA/CE approved photoplethysmography device placed on the finger. All data were captured simultaneously and at rest. The MAX30101 sensor was more accurate in measuring heart rate, blood oxygen saturation and respiration rate compared to the MAX30102. Additionally, we found that the red wavelength was best for measuring heart rate and the green wavelength was best for respiration rate.
No takes yet. Share an insight, caveat, or question.
Longmore et al. (2019) reported a cross-sectional. Tri-wavelength MAX30101 sensor vs. Bi-wavelength MAX30102 sensor and FDA-approved finger device was evaluated on Accuracy in measuring heart rate, blood oxygen saturation, and respiration rate. The tri-wavelength MAX30101 sensor placed on the forehead was more accurate in measuring heart rate, blood oxygen saturation, and respiration rate compared to the bi-wavelength MAX30102 sensor.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: