The article presents the design and testing results of a photoplethysmographic signal sensor intended for use in everyday activities.The design of the sensor involved a commercial DCM03 sensor based on the refl ectance method and a fully integrated analog, front-end AFE4490.Determination of the photoplethysmographic signal and calculation of heart rate was based on appropriate data processing algorithms including digital fi lters, signal segmentation, moving average and variable baseline.Measurements were carried out in diff erent sensor placement locations as well as during specifi c body motions.Detailed analysis of the recorded signal was performed using the proposed processing methods and the fast Fourier transform technique.Based on our results, it is possible to measure HR values and determine the photoplethysmographic curve under dynamic conditions.On the other hand measurement of blood oxygen saturation is subject to signifi cant limitations.