Key result
During intensive vasomotion, the oscillometric finger blood pressure monitor significantly overestimated mean blood pressure compared to Finapres (difference 6.7 mm Hg, SD 2.0).
Why the study?
Does the oscillometric finger blood pressure monitor differ from Finapres in measuring mean blood pressure during intensive vasomotion in healthy volunteers?
Observational (n=6)
Does the oscillometric finger blood pressure monitor differ from Finapres in measuring mean blood pressure during intensive vasomotion in healthy volunteers?
The oscillometric finger blood pressure monitor tends to overestimate mean blood pressure compared to Finapres during peripheral vasoconstriction.
Answers are grounded in this paper and cite their sources.
Warrants caution interpreting oscillometric finger BP during vasomotion; leaves open validation in broader clinical populations.
Two different techniques for non-invasive beat-to-beat finger arterial blood pressure monitoring are compared in six healthy volunteers during local hand heating from 21 to 38 degrees C. The degree of peripheral vasoconstriction was established by recording the thumb pulp skin blood flow with a laser Doppler instrument. For time episodes without vasoconstriction no systematic difference in the readings of beat-to-beat mean blood pressure of the two monitors was found (the oscillometric device UT9201 minus Finapres difference was 0.3 mm Hg, SD 0.3). For the episodes with vasoconstriction the difference was statistically significant (6.7 mm Hg, SD 2.0). The oscillometric device minus Finapres difference and the laser Doppler signal were found to be inversely correlated, the correlation coefficient varying from -0.28 to -0.67. A disagreement between the readings of the instruments during intensive vasomotion is assumed to be caused mainly by the tendency of the oscillometric method to overestimate the finger mean blood pressure under the condition of peripheral vasoconstriction.
No takes yet. Share an insight, caveat, or question.
Katrin Raamat (2000) conducted an observational in Healthy volunteers (n=6). Oscillometric finger blood pressure monitor (UT9201) vs. Finapres was evaluated on Difference in beat-to-beat mean blood pressure readings during episodes with vasoconstriction (MD 6.7 mm Hg). During intensive vasomotion, the oscillometric finger blood pressure monitor significantly overestimated mean blood pressure compared to Finapres (difference 6.7 mm Hg, SD 2.0).