Why the study?
Does the modified oscillometric method (UT9201) agree with the Finapres technique in tracking short-term finger blood pressure changes in healthy volunteers?
Does the modified oscillometric method (UT9201) agree with the Finapres technique in tracking short-term finger blood pressure changes in healthy volunteers?
The UT9201 modified oscillometric monitor tracks short-term MAP rises similarly to the established Finapres technique, though with slight differences during MAP drops.
Answers are grounded in this paper and cite their sources.
UT9201 tracks MAP rises similarly to Finapres but differs on drops; hypothesis-generating for continuous monitoring validation.
Blood pressure recordings obtained by the volume clamp method (Finapres monitor, Ohmeda, USA) were compared with measurements by using a novel modification of the oscillometric method (UT9201 beat-to-beat monitor, University of Tartu, Estonia). Step changes in total peripheral resistance were introduced in 11 healthy volunteers by rapid inflation to the suprasystolic pressure of bilateral thigh cuffs during a rhythmical exercise with the quadriceps muscles and simultaneously performed handgrip compression. The provoked short-term changes in mean arterial pressure (MAP) were similarly tracked by both devices. The agreement between the methods was better for MAP rise than drop. The group-averaged difference UT9201 minus Finapres did not exceed 1.2 mmHg (NS, p = 0.46) after cuff successive inflation but yielded 3.8 mmHg (significant, p < 0.05) after cuff successive deflation.
No takes yet. Share an insight, caveat, or question.
Raamat et al. (2001) studied this question.