Key result
Multislice Joint T1-T2 enabled simultaneous whole left-ventricular T1 and T2 mapping during free breathing in 1.4 minutes, yielding T1 values of 1373 ms vs 1564 ms for standard acquisition.
Why the study?
To develop a free-breathing sequence for simultaneous measurement of myocardial T1 and T2 across multiple slices to achieve whole left-ventricular coverage.
Answers are grounded in this paper and cite their sources.
The Multislice Joint T1-T2 sequence enables simultaneous, whole left-ventricular T1 and T2 mapping during free breathing in under 2 minutes.
Purpose To develop a free‐breathing sequence, that is, Multislice Joint T 1 ‐T 2 , for simultaneous measurement of myocardial T 1 and T 2 for multiple slices to achieve whole left‐ventricular coverage. Methods Multislice Joint T 1 ‐T 2 adopts slice‐interleaved acquisition to collect 10 single‐shot electrocardiogram‐triggered images for each slice prepared by saturation and T 2 preparation to simultaneously estimate myocardial T 1 and T 2 and achieve whole left‐ventricular coverage. Prospective slice‐tracking using a respiratory navigator and retrospective image registration are used to reduce through‐plane and in‐plane motion, respectively. Multislice Joint T 1 ‐T 2 was validated through numerical simulations and phantom and in vivo experiments, and compared with saturation‐recovery single‐shot acquisition and T 2 ‐prepared balanced Steady‐State Free Precession (T 2 ‐prep SSFP) sequences. Results Phantom T 1 and T 2 from Multislice Joint T 1 ‐T 2 had good accuracy and precision, and were insensitive to heart rate. Multislice Joint T 1 ‐T 2 yielded T 1 and T 2 maps of nine left‐ventricular slices in 1.4 minutes. The mean left‐ventricular T 1 difference between saturation‐recovery single‐shot acquisition and Multislice Joint T 1 ‐T 2 across healthy subjects and patients was 191 ms (1564 ± 60 ms versus 1373 ± 50 ms; P < .05) and 111 ms (1535 ± 49 ms vs 1423 ± 49 ms; P < .05), respectively. The mean difference in left‐ventricular T 2 between T 2 ‐prep SSFP and Multislice Joint T 1 ‐T 2 across healthy subjects and patients was −6.3 ms (42.4 ± 1.4 ms vs 48.7 ± 2.5; P < .05) and −5.7 ms (41.6 ± 2.5 ms vs 47.3 ± 2.7; P < .05), respectively. Conclusion Multislice Joint T 1 ‐T 2 enables quantification of whole left‐ventricular T 1 and T 2 during free breathing within a clinically feasible scan time of less than 2 minutes.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2020) studied this question. Multislice Joint T1-T2 sequence vs. Saturation-recovery single-shot acquisition and T2-prep SSFP sequences was evaluated on Mean left-ventricular T1 in healthy subjects (MD 191 ms, p=<.05). Multislice Joint T1-T2 enabled simultaneous whole left-ventricular T1 and T2 mapping during free breathing in 1.4 minutes, yielding T1 values of 1373 ms vs 1564 ms for standard acquisition.
Synapse has enriched 10 related papers. Consider them for comparative context: