Key result
Severe aortic valve calcification deforms the Evolut FX+ frame, narrowing the coronary access window.
Why the study?
Frame deformation of self-expanding valves has been reported in valve-in-valve procedures, but deformation involving a coronary access window of the Evolut FX+ in native aortic valve cases had not been reported.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
This case highlights the risk of severe frame deformation of the Evolut FX+ system's coronary access window when deployed in heavily calcified native aortic valves.
Objective: The Evolut FX+ system features 3 large coronary access windows to facilitate coronary cannulation. Frame deformation of self-expanding valves has been reported in valve-in-valve procedures. However, to our knowledge, this is the first report of severe deformation involving a coronary access window in a native aortic valve case.
No takes yet. Share an insight, caveat, or question.
Hasegawa et al. (2026) conducted a case report in Severe aortic stenosis (n=1). Evolut FX+ system (TAVI) was evaluated on Frame deformation. Severe calcification in a native aortic valve caused significant localized frame deformation of the Evolut FX+ system, leading to marked narrowing of a large coronary access window.
Synapse has enriched 10 related papers. Consider them for comparative context: