Key result
Balloon post-dilatation resolves 34 mm Evolut R frame infolding caused by loss of nitinol memory.
Case Report
Alerts to rare infolding risk with 34 mm Evolut R; leaves open nitinol memory preservation strategies in TAVI.
Figure 1. Computed tomography of native aortic annulus and serial fluoroscopic images of transcatheter aortic valve implantation procedure. A) Computed tomography of aortic annulus shows calcium at clock positions 2 and 6. B) In vitro fluoroscopic confirmation of the appropriately loaded 34 mm Evolut R valve system. C) Infolding of the valve frame is shown in between the arrows (dotted dark blue area in the cartoon inset). D) Underexpanded valve frame after complete release of the valve (dotted dark blue area in the cartoon inset). E) Balloon dilatation of the underexpanded valve frame with a 25 mm × 5 cm Z-MED II balloon. F) ...
No takes yet. Share an insight, caveat, or question.
Vasu et al. (2020) conducted a case report in Transcatheter aortic valve implantation complication. Transcatheter aortic valve implantation (Evolut R valve system) was evaluated. Infolding of a 34 mm Evolut R self-expanding bioprosthesis occurred due to loss of nitinol memory during transcatheter aortic valve implantation, requiring balloon dilatation.