Aerosol Delivery With a Vibrating Mesh Nebulizer Across Tidal Volume-Based Pediatric Invasive Ventilation Models: An In Vitro Evaluation

View Full Paper
Ask AI
Bookmark
Share

Authors

SMShinya MiuraYIYoshihiro IgarashiKSKenta Shono

Discussion

Loading...

Member takes

Overview

In vitro experiment evaluates aerosol delivery efficacy in pediatric ventilation models, indicating tidal volume's importance.

Key Points

  • Evaluate how tidal volume-based ventilation models affect aerosol delivery efficacy using a vibrating mesh nebulizer in pediatrics.
  • Conducted in vitro experiments with six pediatric ventilation models at tidal volumes of 40–400 mL.
  • Nebulized salbutamol using a vibrating mesh nebulizer positioned on the dry side of a heated humidifier.
  • Quantified inhaled mass fraction using a spectrophotometer to measure drug collected on a filter adjacent to the endotracheal tube.
  • Inhaled mass fraction increased with tidal volume: 3.5% at 40 mL, 7.9% at 56 mL, 15.7% at 120 mL, 19.4% at 160 mL, and 21.3% at 400 mL (p < 0.001 vs. baseline).
  • No significant differences in aerosol delivery were found with variations in other ventilatory settings or endotracheal tube sizes.