Answers are grounded in this paper and cite their sources.
post-operative radiographs by two independent reviewers and compared against coronal alignment angles from the CAN. The primary outcome was the mean absolute difference of femoral and tibial varus/valgus angles between radiograph and intra-operative device measurements. Bland-Altman plots were used to assess agreement between the methods and statistically analyze potential systematic bias. The mean absolute differences between navigation-guided cut measurements and post-operative radiographs were 1.16 ± 1.03° and 1.76 ± 1.38° for femoral and tibial alignment respectively. About 88% of coronal measurements were within ±3°, while 99% were within ±5°. Bland-Altman analysis demonstrated a bias between CAN and radiographic measurements with CAN values averaging 0.52° (95% CI: 0.11°-0.93°) less than their paired radiographic measurements. This study demonstrated the ability of an optical imageless navigation system to measure, on average, femoral and tibial coronal cuts to within 2.0° of post-operative radiographic measurements in a clinical setting.
No takes yet. Share an insight, caveat, or question.
Winberg et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: