Key result
Overnight pulse oximetry using a 3% oxygen desaturation index >7.5/h had 97% sensitivity and an AUC of 0.92 for detecting sleep-disordered breathing in patients with chronic heart failure.
Why the study?
Do overnight pulse oximetry and heart rate variability analysis accurately screen for sleep-disordered breathing in patients with chronic heart failure?
Cross-Sectional (n=180)
Do overnight pulse oximetry and heart rate variability analysis accurately screen for sleep-disordered breathing in patients with chronic heart failure?
Overnight pulse oximetry is a highly sensitive screening tool to rule out sleep-disordered breathing in chronic heart failure, whereas heart rate variability analysis is not useful.
May support pulse oximetry screening for sleep-disordered breathing in chronic heart failure; leaves open prospective validation before clinical adoption.
BACKGROUND: Sleep-disordered breathing (SDB) is under diagnosed in chronic heart failure (CHF). Screening with simple monitors may increase detection of SDB in a cardiology setting. This study aimed to evaluate the accuracy of heart rate variability analysis and overnight pulse oximetry for diagnosis of SDB in patients with CHF. METHODS: 180 patients with CHF underwent simultaneous polysomnography, ambulatory electrocardiography and wrist-worn overnight pulse oximetry. SDB was defined as an apnoea-hypopnoea index ≥15/h. To identify SDB from the screening tests, the per cent very low frequency increment (%VLFI) component of heart rate variability was measured with a pre-specified cutoff ≥2.23%, and the 3% oxygen desaturation index was measured with a pre-specified cutoff >7.5 desaturations/h. RESULTS: 173 patients with CHF had adequate sleep study data; SDB occurred in 77 (45%) patients. Heart rate variability was measurable in 78 (45%) patients with area under the %VLFI receiver operating characteristic curve of 0.50. At the ≥2.23% cutoff, %VLFI sensitivity was 58% and specificity was 48%. The 3% oxygen desaturation index was measurable in 171 (99%) patients with area under the curve of 0.92. At the pre-specified cutoff of >7.5 desaturations/h, the 3% oxygen desaturation index had a sensitivity of 97%, specificity of 32%, negative likelihood ratio of 0.08 and positive likelihood ratio of 1.42. Diagnostic accuracy was increased using a cutoff of 12.5 desaturations/h, with sensitivity of 93% and specificity of 73%. CONCLUSIONS: The high sensitivity and low negative likelihood ratio of the 3% oxygen desaturation index indicates that pulse oximetry would be of use as a simple screening test to rule out SDB in patients with CHF in a cardiology setting. The %VLFI component of heart rate variability is not suitable for detection of SDB in CHF.
No takes yet. Share an insight, caveat, or question.
Ward et al. (2012) conducted a cross-sectional in chronic heart failure (n=180). Overnight pulse oximetry and heart rate variability analysis vs. Polysomnography was evaluated on Diagnosis of sleep-disordered breathing (apnoea-hypopnoea index ≥15/h) (AUC 0.92). Overnight pulse oximetry using a 3% oxygen desaturation index >7.5/h had 97% sensitivity and an AUC of 0.92 for detecting sleep-disordered breathing in patients with chronic heart failure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: