Key result
Magnetic resonance imaging planimetry of the proximal vena contracta accurately quantified the true orifice area in an idealized, constant-flow model.
Why the study?
Does MRI planimetry of the proximal vena contracta accurately measure the orifice area in an in vitro constant-flow model of stenotic valves?
Does MRI planimetry of the proximal vena contracta accurately measure the orifice area in an in vitro constant-flow model of stenotic valves?
MRI planimetry of the proximal vena contracta accurately measures the orifice area in an in vitro constant-flow model, suggesting potential utility for assessing stenotic heart valves.
Answers are grounded in this paper and cite their sources.
Accurate in idealized models; leaves open clinical translation and in vivo validation for stenotic valve assessment.
Exact determination of the orifice area in stenotic valve disease is essential to guide therapy. With the standard imaging methods, the orifice area has to be calculated by empirically-derived formulas, which may be susceptible to changes in hemodynamic status, leading to wrong therapeutic decisions. We tested the ability of magnetic resonance imaging (MRI) to quantify the orifice area by planimetry of the proximal vena contracta (PVC) in an idealized, constant-flow model in a 1.0-T tomograph (Siemens Magnetom Expert). There was a close correlation of the PVC area in MRI and the true orifice area (r2 = 0.985), and agreement of the methods as measured by the Bland-Altman test was significant with a low systematic error of -0.02 cm2. We conclude that MRI planimetry of the PVC in stenotic valves in a constant-flow model may be used to reliably measure the orifice area.
No takes yet. Share an insight, caveat, or question.
Strohm et al. (2001) studied Stenotic valve disease. Magnetic resonance imaging (MRI) planimetry of the proximal vena contracta vs. True orifice area was evaluated on Correlation of the PVC area in MRI and the true orifice area (r2 = 0.985). Magnetic resonance imaging planimetry of the proximal vena contracta accurately quantified the true orifice area in an idealized, constant-flow model.
Synapse has enriched 10 related papers. Consider them for comparative context: